High frequency electrosurgical equipment suitable to deliver current for monopolar cut and coagulated cut (with different levels of coagulation), bipolar cut and coagulation.
TWIN OPERATIVE ACTIVITY
- two output channels for monopolar cutting
- two independent setting and output channels for the most frequent monopolar coagulation
CONTROLLED CUTTING PROCEDURES
- enhanced cutting current for eschar free incision
- cutting current with selectable coagulating property for bleeding controlled dissection
- low capacitance coupling by slow BLEND current for endoscopic procedures
TWIN CONTROLLED COAGULATION PROCEDURES
- SPEEDY coagulation and DEEP coagulation available to two operators
- SPRAY coagulation by hand switch or foot switch controls
EXTENSIVE BIPOLAR FEATURES
- bipolar cutting current with selectable bleeding control
- slow BLEND cutting current
- automatic Start/Stop coagulation functions
SURGEON AND PATIENT TOTAL SAFETY·
- acceptability of contact impedance between patient body and return electrode continuously monitored by Skin Plate Electronic Control (SPEC)
- H.F. leakage currents continuously monitored
- mains fluctuation automatically compensated before delivery of HF power
PERSONALIZED PROCEDURES
- independent choice of coagulation currents and power levels by each of two operators
- up to 10 procedures can be configured and re-called
EXTENSION DRIVER AND INTERLOCK SIGNALS
- driven output connection to external Argon Plasma unit and Smoke Evacuator unit
AUTOMATIC SELF DIAGNOSIS
- automatic diagnostic control of the complete functionality of electronic unit and connected accessories with coded report of failures
SUITABLE IN:
- OUTPATIENT SURGERY
- GENERAL SURGERY
- PAEDIATRIC SURGERY
- PLASTIC SURGERY
- THORAX SURGERY
- VASCULAR SURGERY
- DERMATOLOGY
- ENDOSCOPY
- GASTROENTEROLOGY
- GYNAECOLOGY
- NEUROSURGERY
- ORTHOPEDICS
- OTORHINOLARINGOLOGY
- PNEUMOLOGY
- UROLOGY
- VETERINARY
Technical Specifications
• Maximum out power monopolar pure CUT: 300 W - 300 O
• Maximum out power monopolar cut-coag CUT80%: 250 W - 300 O
• Maximum out power monopolar cut-coag CUT60%: 200 W - 300 O
• Maximum out power monopolar cut-coag CUT40%: 150 W - 300 O
• Maximum out power monopolar cut ENHANCED: 200 W - 500 O
• Maximum out power monopolar cut-coag BLEND: 200 W - 300 O
• Maximum out power monopolar coag SPEEDY: 100 W - 500 O
• Maximum out power monopolar coag DEEP: 100 W - 200 O
• Maximum out power monopolar coag SPRAY 70 W - 2000 O
• Maximum out power monopolar coag SPRAY ARGON: 70 W - 2000 O
• Maximum out power bipolar CUT 95 W - 150 O
• Maximum out power bipolar cut-coag CUT80%: 95 W - 150 O
• Maximum out power bipolar cut-coag CUT60%: 95 W - 150 O
• Maximum out power bipolar cut-coag CUT40%: 60 W - 150 O
• Maximum out power bipolar cut-coag BLEND: 90 W - 150 O
• Maximum out power bipolar cut-coag BLEND: 70 W - 100 O
• Working frequency monopolar/bipolar: 425 kHz / 525 kHz
• Patient circuit: F
• Selectable input voltage: 115-230 Vac
• Mains frequency: 50-60 Hz
• Electrical input voltage: 1000 VA
• Size WxHxD mm: 470 x 150 x 400
• Weight: 17.5 kgs
• Controls: Patient/Plate circuit monitoring, Output power monitoring, Self check control
• Safety: EN60601-1, EN60601-1-2, EN60601-2-2
• Electrical Class: I CF, MDD 93/42/EC - Class: II b
Standard accessories
• 1 power supply cord 5 m
• 1 two-button handle (autoclavable code 30539)
• 5 two-button handles (cold sterilizable code 30551)
• 1 cable for flat plate (30559)
• 2 HP double water-proof foot switches
• 3 short, general surgery blade electrodes 7 cm
• 3 long, general surgery blade electrodes 16 cm
• 3 short, general surgery ball electrodes Ø 4 mm, 6 cm
• 1 electrode brush