PAVO Vital Sign Monitor
Features
• 8” color TFT LCD Screen (touch screen optional)
• Portable, lightweight and sturdy design
• Flexible parameters configuration for different clinical environments
• Rechargeable Li-ion Battery (up to 12 hours uninterruptable work)
• Big font and font color display setting
• Spot-check and continuous monitoring mode
• Selectable for Adult, Pediatric and Neonatal patients
• Wired/Wireless CMS, support HL7 protocol to HIS
• Barcode scanner support • Thermal recorder support
• Graphical & tabular trend review
• 48 hours holographic wave review for each patient (stored in SD card)
Display
8” color TFT LCD Screen, resolution: 800 x 600
ECG Lead type 3-lead: I, II, III--5-lead: I, II, III, aVR, aVL, aVF, V
Display sensitivity:
2.5 mm/mV (×0.25), 5 mm/mV (×0.5),10 mm/mV (×1.0), 20 mm/mV (×2.0)
Wave sweep speed: 6.25 mm/s, 12.5 mm/s,
25 mm/s, 50 mm/s
Bandwidth
Diagnostic mode: 0.05Hz~100Hz
Monitor mode: 0.5Hz~40Hz
Surgery mode: 1Hz~20Hz
Strong filter mode: 5Hz~20Hz CMRR
>100dB
Notch: 50/60 Hz notch filter can be set to on or off
Differential input impedance >5 MΩ
Electrode polarization voltage range: ±400 mV
Baseline recovery time <3 s after defibrillation (in monitor and surgery mode)
Calibration signal: 1 mV (peak - peak),
accuracy ±3%
RESP
Measurement method: Thoracic electrical
bioimpedance
Rate: 0 – 150 bpm
Measuring lead: Lead I, II
Wave gain: ×0.25, ×0.5, ×1, ×2
Respiratory impedance range: 0.5-5 Ω
Baseline impedance: 500-4000 Ω
Gain: 10 grades
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s
TEMP
Measurement method: Thermistor
Measuring range: 5~50 °C (41~122 °F)
Resolution: 0.1 °C
Measurement accuracy: ±0.1 °C
Recorder (optional)
Built-in, Thermal dot array
Horizontal resolution: 16 dots/mm (25 mm/spaper speed)
Vertical resolution: 8 dots/mm
Paper speed: 25 mm/s, 50 mm/s
Number of waveform channels: 3